26
Internal Calibration (ICal)
There is no need to recalibrate the meter for milliamps when first received
from the factory.
The meter is factory calibrated for milliamps prior to shipment. The
calibration equipment is certified to NIST standards.
The internal calibration allows the user to scale the meter without applying a
signal. The use of a calibrated signal source is necessary to perform the internal
calibration of the meter. Check calibration of the meter at least every 12 months.
Notes:
The signal source must have a full-scale accuracy of 0.01% or better between
4 and 20 mA in order to maintain the specified accuracy of the meter.
Allow the meter to warm up for at least 15 minutes before performing the
internal calibration procedure.
The Internal calibration menu is part of the Advanced features menu.
Press and hold the Menu button for 5 seconds to enter the Advanced features
menu. Press the Up arrow button to scroll to the Internal Calibration menu (ICAL)
and press Enter.
The meter displays 4.00 mA. Apply a 4.00 mA signal and press Enter. The
display flashes for a moment while the meter is accepting the signal.
After the signal is accepted, the meter displays 20.00 mA. Apply a 20.00 mA
signal and press Enter. The display flashes for a moment while the meter is
accepting the signal.
Error Message (Err)
An error message indicates that the calibration process was not successful. After
the error message is displayed, the meter will revert to input 2 calibration
settings. The error message might be caused by inadvertently leaving the signal
at the previous level or not maintaining a 0.40 mA minimum span. Press the
Menu button to cancel the current calibration process if necessary.